Purpose
The
UMBRELLA kit for photodynamic therapy and fluorescent diagnostics
is intended for visual diagnostics and treatment by a method of
photodynamic therapy of colonies of pathological and oncological cells.
The
kit is applied independently or together with other therapeutic or surgical
laser instruments, laser spectral evaluators and diverse diagnostic instruments
in structure of the cabinets of photodynamic therapy in clinics, hospitals and
research and development medical establishments.
The
complex reputes operation with different types of the PDT-drugs, first of all
with high-performance and low-toxic ones, for example, with the Radachlorin
photosensitizer.
Device and principle of the operation
The
kit is executed on the basis of a matrix from semiconductor light emitting
diodes of heightened brightness and specialized video camera with a specially
designed light filter providing authentic registration of a fluorescence. The
kit reputes usage of standard means of a signal display of the video map of the
camera allowed for medical practice.
The
kit powers up a matrix and video camera hardened on supports permitting to
direct radiating of a matrix and a field of supervising of the camera on the
same zone, for example, surface segment of a skin. A matrix and video camera
aggregate by cables to a separate power supply unit, through last the cue on the
screen monitor files also.
The
matrix contains group (more than 1000) light emitting diodes built - on
close-packed on the shaped round surface permitting to focalize (in definite
limits) total radiating of collection of diodes of a matrix. The matrix is
supplied with cooling vanes. The video camera is executed on the basis of
standard vehicle KRS-900 in the optical channel which one the filtering of a
visible portion of the spectrum is stipulated, except for area of a fluorescence,
for example, 670-680 nm.
The
matrix is used for beaming a zone of a pathology previously sensitized by the
PDT-drug. Thus the wave length of radiating of light emitting diodes of a matrix
should coincide absorption band of the drug. The beaming serves for creation of
a necrosis, thus there is an accompanying fluorescence in an exposure zone of
the drug accrued in a tissue.
The
supervising of an irradiated surface segment with the help of the video camera
allows to register a zone of a fluorescence by the way visually of brighter
pattern indemnification a background of beaming of a matrix. The zone of a
fluorescence meets to a zone of a pathology with heightened accruing of the
pdt-drug a procedure of supervising of a fluorescence of a zone of a pathology
is diagnostic. It can precede, accompany therapeutic beaming or be executed
separately. In the latter case radiation intensity can be reduced by a matrix (power
density and energy) concerning therapeutic significances.
Due
to a possibility of multiple repetition of a procedure at usage of the
Radachlorin photosensitizer (feature of the drug favourably distinguishing it
from other photosensitizers) is possible to watch process of diagnostics -
therapy in dynamics.
Characteristics
- Radiation
condition of a matrix - continuous
- Radiation
wavelength of a matrix - 650-670 nm
- A
power density of working radiating of a matrix in a focal plan - 30-60 mW/sm2
- Consumed
electric power of the vehicle - no more than 20 W
- The
floor space of a spot of beaming in a focal plan - 40 sm2
- Focal
spacing interval - 20-40 sm
- Overall
dimensions of a matrix no more – 300 x 300 mm
- A
definition of an oilpainting of a bloom on the shield of a black-and-white
monitor with resolution - no more than 2 mms
- A
transient time of an operating duty - no more than 5 min.
- Time
of continuous operation - till 6 hours
- A
supply voltage - 220 In (+/- 10%)
- Frequency
of a web - 50 Hz
- Consumed
power - no more than 250 VA
- Bulk
of kit - no more than 12 kgs
- An
average operating time of the vehicle on refusal not less than 1500 hours,
average due date of a service of the vehicle - 5 years
The
electrosafety of the vehicle conforms the requirements the GOST R 50267.0 and is
executed on the class 2 and type of protection BF. The vehicle is thermostable
at operation to effect of temperature of ambient air from 10 up to 35oя.
List of standard equipment
The
UMBRELLA kit for photodynamic therapy and fluorescent diagnostics
RDFA.944420.001* in structure:
- Power
supply unit of kit (RDFA.944421.001) - 1 piece
- The
video camera KRS-900 (RDFA.311528.001)** - 1 piece
- LED
matrix (RDFA.943635.001) - 1 piece
- Power
cord AWM E69326 - 1 piece
- The
passport TO and IE (RDFA.944420.001PS) - 1 piece
- Bag
for a storage and transportation of kit - 1 piece
- Support
LPD24.MN*** - 2 pieces
- Glasses
protective**** - 1 piece
* - The video monitor does not enter into a complete set of delivery, is if
necessary delivered to the customer the separate payment.
** - The video camera does not enter into the simplified set of delivery.
*** - Acquisition by supports to be made separately, as agreed with the customer.
**** - Acquisition by shielding spectacles to be made separately, as agreed with
the customer. The complete set by glasses of other pattern with the similar
performances is possible.
For
usage of the UMBRELLA kit for photodynamic therapy and fluorescent diagnostics
with the personal computer, the problem software permitting maximum effectively
is developed to watch dynamics of treatment and diagnostics of the patients.
By the order to be made and the
simplified version of a complex is delivered, which one does not include the
video camera.